SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 433
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO
<character_data>
This command is used to trigger Auto Gating and turning on or off the
Perpetual Gating for the selected gate.
Syntax
Parameters
NOTE
This command is only applicable when N192x or E932x sensor is present and trigger
acquisition mode is selected.
Space
ON
?
1
2
4
3
OFF
:AUTO
SENS
1
SWE
SENS2
:
ONCE
Item Description/Default Range of Values
character_data The status of Auto Gating and Perpetual
Gating.
ONCE: To turn on Auto Gating
ON/OFF: To turn on/off Perpetual Gating
OFF
ON
ONCE