10 SENSe Subsystem
434 N1911A/1912A P-Series Power Meters Programming Guide
Example
Reset Condition
On reset, Perpetual Gating will be disabled.
Query
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO?
The query returns the current setting of perpetual gating(0 or 1).
• 1 is returned if the perpetual gating is turned on
• 0 is returned if the perpetual gating is turned off
Query Example
SENS:SWE2:AUTO ON This command turns on Channel A Gate 2
Perpetual Gating.
SENS2:SWE3:AUTO OFF This command turns off Channel B Gate 3
Perpetual Gating.
SENS2:SWE4:AUTO ONCE This command triggers Auto Gating for
Channel B Gate 4.
SENS2:SWE:AUTO? The query returns the current setting of
Perpetual Gating for Gate 1 Channel B.