PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 317
PSTatistic:CCDF:MARKer[1]|2:X <numeric_value>
This command is used to set the current marker X- axis position at the
selected trace.
Syntax
Parameters
1
The marker will be placed at the point nearest to the specified X-axis position if that particular X
value is not available.
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
• Expanded window is enabled (not full CCDF screen window)
PST
:CCDF
2
: X
1
:MARK
Space
numeric_value
Item Description/Default Range of Values
numeric_value
The current marker X-axis position
1
.
• Maximum Value: 50 dB. The value will be
set to 0 if negative value is inserted.
<
50