SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 463
SENSe[1]|2:TRACe:Y:SCALe:PDIV <numeric_value>
This command is used to set the Y Scale value (per division) for the
selected trace.
The Y Scale value set by this SCPI command is dependant on the current
Y- axis unit, which can be set using SENS:TRAC:UNIT command.
Syntax
Parameters
NOTE
This command is only applicable when N192x or E932x sensor is present and trigger
acquisition mode is selected.
Item Description/Default Range of Values
numeric_value The numeric value for Y-axis scale. 0.001 dB to10 dB
1 nWatt to10 MWatt
1
2
Space
?
:Y
SENS
:TRAC
:SCAL
:PDIV
numeric_value