10 SENSe Subsystem
462 N1911A/1912A P-Series Power Meters Programming Guide
Reset Condition
On reset, X Start = 0 s; X Scale = 10 µs.
Query
SENSe[1]|2:TRACe:X:SCALe:PDIV?
The query returns the current scale setting of X- axis in numerical value.
Query Example
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
• If channel is not in trigger acquisition mode, error –221 "Settings
conflict" occurs.
• If limits of the values keyed in are exceeded, error –222 "Data out of
range; upper (or lower) limit exceeded; no change" occurs.
SENS:TRAC:X:SCAL:PDIV? This command queries the Channel A
current X- axis scale setting in numerical
value.