Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 43
Using STATus
If using GPIB, you can use the STATus subsystem to generate an SRQ to
interrupt your program when a limit failure occurs. This is a more
efficient method than using SENSe or CALCulate, since you do not need
to check the limit failures after every power measurement.
Refer to “Status Reporting” on page 50 and “STATus Subsystem” on
page 465 for further information.
NOTE
If TRIGger:DELay:AUTO is set to ON, then the number of failures returned by
SENSe:LIMit:FCOunt? or CALCulate[1|2]:LIMit:FCOunt? is affected by the
current filter settings.