PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 351
PSTatistic[1]|2:CCDF:TRACe:POWer:AVERage?
This command is used to retrieve average power value of Channel A or
Channel B trace.
Syntax
Example
Error Messages
• If no power sensor is connected, error –241 "Hardware missing" occurs.
• If sensor/sensors connected are not P- Series sensors, error –241
"Hardware missing" occurs.
• If measurement speed setting is FAST, error –221 "Settings conflict"
occurs.
• If command executed in other window besides CCDF window, error
–221 "Settings conflict: Requires CCDF window" occurs.
This command is only applicable when P-Series sensors are present and the following
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
:POW
PST
2
1
:CCDF
:TRAC
?
:AVER
PST:CCDF:TRAC:POW:AVER? This command returns the average power
value for Channel A trace.