10 SENSe Subsystem
392 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect]
<string>
This command enters the name of the sensor calibration table or
frequency dependent offset table which is to be used. The CSET1 command
selects the sensor calibration table and the CSET2 command selects the
frequency dependent offset table. The calibration factor is interpolated
from the table using the setting for [SENSe[1]]|SENSe2:FREQuency.
Syntax
Parameters
If [SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe is set to OFF, the
selected sensor calibration table or frequency offset table is not being used.
Item Description/Default Range of Values
string String data representing a sensor
calibration table, or frequency
dependent offset table name.
Any existing table name
(Existing table names
can be listed using
MEMory:CATalog:TABle?).
SENS
1
:CSET
CORR
SENS2
:
?
:SEL
string
Space
1
:CSET2