SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 393
Example
Reset Condition
On reset the selected table is not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table
is selected an empty string is returned.
Query Example
Error Messages
•If <string> is not valid, error –224, “Illegal parameter value” occurs.
• If a table called <string> does not exist, error –256, “File name not
found” occurs.
• When a sensor calibration table is selected, the power meter verifies
that the number of calibration points defined is one more than the
number of frequency points defined. When a frequency dependent
CORR:CSET1 ‘PW1’ This command enters the name of the
sensor calibration table which is to be
used on Channel A.
CORR:CSET1? This command queries the sensor
calibration table currently used for
Channel A.