12 SYSTem Subsystem
506 N1911A/1912A P-Series Power Meters Programming Guide
SYSTem:COMMunicate:LAN:DNAMe <character_data>
This command sets the domain name for the power meter.
Syntax
Parameters
Example
Query
SYSTem:COMMunicate:LAN:DNAM?
The query returns the current setting of the LAN domain name.
SYST
:COMM
:LAN
:DNAM
Space
?
character_data
Item Description Range of Values
character_data Character values of up to 16 characters Maximum of 16
characters
SYST:COMM:LAN:DNAM
‘myco.com’
This command sets the hostname to
myco.com.