13 TRACe Subsystem
638 N1911A/1912A P-Series Power Meters Programming Guide
TRACe[1]|2:STATe <boolean>
This command enables or disables trace capture for the specified channel.
Syntax
Example
Reset Condition
On reset trace capture is set to OFF.
Query
TRACe[1]|2:STATe?
The query command enters a 1 or 0 into the output buffer indicating
whether or not trace capture is enabled or disabled.
NOTE
This command does not allow the setting set to ON when either measurement channel (for
dual channel) is configured to initiate trigger buffering.
TRAC
0|OFF
1|ON
Space
?
:STAT
1
2
TRAC2:STAT 1 This command enables trace capture for
Channel B.