EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #666 background imageLoading...
Page #666 background image
13 TRACe Subsystem
638 N1911A/1912A P-Series Power Meters Programming Guide
TRACe[1]|2:STATe <boolean>
This command enables or disables trace capture for the specified channel.
Syntax
Example
Reset Condition
On reset trace capture is set to OFF.
Query
TRACe[1]|2:STATe?
The query command enters a 1 or 0 into the output buffer indicating
whether or not trace capture is enabled or disabled.
NOTE
This command does not allow the setting set to ON when either measurement channel (for
dual channel) is configured to initiate trigger buffering.
TRAC
0|OFF
1|ON
Space
?
:STAT
1
2
TRAC2:STAT 1 This command enables trace capture for
Channel B.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals