14 TRIGger Subsystem
656 N1911A/1912A P-Series Power Meters Programming Guide
INITiate[:IMMediate]:ALL
This command initiates all trigger sequences.
Syntax
Example
Error Messages
If the power meter is not in the idle state or INITiate:CONTinuous is ON,
error –213, “INIT ignored” occurs.
INIT:IMM:ALL This command initiates all trigger
sequences.
INIT
:IMM
:ALL