TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 655
Reset Condition
On reset (*RST), this command is disabled.
On preset (SYSTem:PRESet) and instrument power- up, this command is
enabled.
Query
INITiate[1]|2:CONTinuous:SEQuence?
The query enters a 1 or 0 into the output buffer.
1 is returned when there is continuous triggering
0 is returned when there is only a single trigger
Query Example
INIT2:CONT:SEQ? This command queries whether Channel B
is set for single or continuous triggering.