PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 355
PSTatistic[1]|2:CCDF:TRACe:POWer:PTAVerage?
This command is used to retrieve peak to average power value of Channel
A or Channel B trace.
Syntax
Example
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P- Series sensors, error –241
"Hardware missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict"
occurs.
If command is executed in other window besides CCDF window, error
–221 "Settings conflict: Requires CCDF window" occurs.
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
:POW
PST
2
1
:CCDF :TRAC
?
:PTAV
PST:CCDF:TRAC:POW:PTAV? This command returns the peak to
average power value of Channel A trace.