MEASurement Commands 2
N1911A/1912A P-Series Power Meters Programming Guide 97
CONFigure [1] |2|3|4 Commands
The CONFigure commands are used on the specified
window/measurement to set:
• The expected power level being measured.
• The resolution of the window/measurement.
• The channel(s) on which the measurement is to be made.
The CONFigure commands do not make the power measurement after
setting the configuration. Use READ?, or alternatively use INITiate
followed by a FETCh? to make the measurement.
The CONFigure command also applies the following defaults to the
channel(s) which are in the specified window (the channel(s) in the
window are specified in the <source list> parameter):
Default Settings Description
INITiate:CONTinuous OFF Sets the power meter to make one trigger cycle
when INITiate is sent.
TRIGger:SOURce IMMediate When TRIG:SOUR is set to BUS or HOLD, sets
the power meter to make the measurement
immediately a trigger is received.
TRIGger:DELay:AUTO ON Enables automatic delay before making the
measurement.
SENSE:AVERage:COUNt:AUTO ON Enables automatic filter length selection.
SENSE:AVERage:STATe ON Enables averaging.