PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 313
PSTatistic:CCDF:MARKer:DELta?
This command is used to retrieve power and probability difference
between marker 2 and marker 1 at any trace the marker is positioned.
Algorithm
Delta = Marker 2 Value - Marker 1 Value
where
Delta is the power difference and probability difference
Syntax
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
• Free-run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen
PST
:CCDF
:MARK
?
:DEL