EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 319
PSTatistic:CCDF:MARKer[1]|2:Y <numeric_value>
This command is used to set the current marker Y- axis position at the
selected trace.
Syntax
Parameter
1
The marker will be placed at the point nearest to the specified Y-axis position if that particular Y
value is not available.
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
PST
:CCDF
2
:Y
1
:MARK
Space
numeric_value
Item Description/Default Range of Values
numeric_value
The current marker Y-axis position
1
.
Minimum Value: 0 %
Maximum Value: 100 %
0 to 100

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals