SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 409
Query Example
Error Messages
If a loss/gain correction value is entered using
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude] while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings
Conflict” occurs. However, the correction value is set but the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command is not
automatically set ON.
•The SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]
command can be used for the 8480 Series power sensor when no
sensor calibration table has been set up.
CORR:GAIN2? This command queries the current setting
of the channel offset on Channel A.