EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #437 background imageLoading...
Page #437 background image
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 409
Query Example
Error Messages
If a loss/gain correction value is entered using
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude] while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings
Conflict” occurs. However, the correction value is set but the
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe command is not
automatically set ON.
•The SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]
command can be used for the 8480 Series power sensor when no
sensor calibration table has been set up.
CORR:GAIN2? This command queries the current setting
of the channel offset on Channel A.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals