4 CALibration Subsystem
216 N1911A/1912A P-Series Power Meters Programming Guide
Reset Condition
On reset, automatic zeroing is disabled.
Query
CALibration[1]|2:ZERO:NORMal:AUTO?
The query always returns a value of 0.
Error Messages
If zeroing was not carried out successfully the error –231, “Data
Questionable; ZERO ERROR” occurs. If you are using a dual channel
power meter, the error message specifies which channel failed zeroing.
If this command is set to ON the error –224, “Illegal parameter value”
occurs.
If there is no sensor connected, or if a sensor other than an E9320 or
P- Series is connected, the error –241, “Hardware missing” occurs.
If an E9320 sensor is connected and is not in NORMAL mode, the
error –221 “Settings conflict” occurs.