PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 345
Syntax
Example
PST
1
:CCDF :TAB
2
?
PST:CCDF:TAB? This command returns the data in CCDF
table: average input power, probabilty at
the average input power, power level at
various predefined probability steps (10
%, 1 %, 0.1 %, 0.01 %, 0.001 % and 0.0001
%), peak to average power ratio and
sample count for Channel A.
PST2:CCDF:TAB? This command returns the data in CCDF
table: average input power, probabilty at
the average input power, power level at
various predefined probability steps (10
%, 1 %, 0.1 %, 0.01 %, 0.001 % and 0.0001
%), peak to average power ratio and
sample count for Channel B.
NOTE
The sample count will always be returned as 100 million samples (100,000,000).