PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 333
Reset Condition
On reset, the CCDF cummulative counts will be set to the default value,
100 M samples.
Query
PSTatistic[1]|2:CCDF:COUNt?
The query returns the current numeric value of the CCDF cummulative
count for the respective channel selected.
Query Example
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
• If channel is not in free- run acquisition mode, error –221 "Settings
conflict" occurs.
• If user inputs a count that is lesser than 100 M or greater than 10 G,
error –222 "Data out of range" occurs. Any input count that is within
range will be normalized and round down to the nearest 100 M.
PST1:CCDF:COUN? This command queries the numeric value
of CCDF cummulative counts for Channel
A.