SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 385
Example
Query
[SENSe[1]]|SENSe2:BUFFer:MTYP?
This query is used to retrieve measurement type settings.
Query Example
On Reset
On *RST, the value is set to AVER
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
If E9320 or N1920 sensor is connected but acquisition mode is in free
run, error –221, “Setting conflict. Invalid acquisition mode” occurs.
If E9320 or N1920 sensor is connected but sensor mode is average,
error –221, “Setting conflict. Invalid acquisition mode” occurs.
If E9320 or N1920 sensor is connected but trigger source is not
external, error- 221 “Setting conflict. Invalid acquisition mode” occurs.
BUFF:MTYP AVER This command sets the measurement type
to average for Channel A.
BUFF:MTYP? This query returns the measurement type
to average for Channel A.