10 SENSe Subsystem
400 N1911A/1912A P-Series Power Meters Programming Guide
Reset Condition
On reset, the duty cycle is set to 1 % (DEF).
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut]
[:MAGNitude]? [MIN|MAX]
The query returns the current setting of the duty cycle or the values
associated with MIN and MAX.
Query Example
Error Messages
If a duty cycle value is entered using
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings
Conflict” occurs. However, the duty cycle value is set but the
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe command is
not automatically set ON.
If this command is used when an E4412A/E4413A power sensor is
connected, the error –310, “System error;Dty Cyc may impair accuracy
with ECP sensor” occurs. If you are using a dual channel power meter
the error message specifies the channel.
If this command is used when a E9320 Series power sensor is set to
NORMal mode, the error –221, “Settings Conflict” occurs.
If this command is used when a P- Series power sensor is connected,
the error –241, “Hardware missing” occurs.
CORR:GAIN3? This command queries the current setting
of the duty cycle for Channel A.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals