10 SENSe Subsystem
420 N1911A/1912A P-Series Power Meters Programming Guide
Example
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STEP?
This query is used to retrieve the number of steps in average or peak
frequency sweep.
Query Example
On Reset
On *RST, the value is set to 0.
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
• If E4410, N8480 (excluding Option CFT), E9300, E9320 or N1920 sensor
is connected but acquisition mode is in free run, error –221, “Setting
conflict. Invalid acquisition mode” occurs.
• If parameter set is lower than 0, error –222, “Data out of range; value
clipped to lower limit” occurs.
FREQ:STEP 10 This command sets frequency sweep with
10 steps for Channel A.
FREQ:STEP? This query returns the number of steps in
frequency sweep for Channel A.