10 SENSe Subsystem
424 N1911A/1912A P-Series Power Meters Programming Guide
Example
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STOP?
This query is used to retrieve stop frequency of the average or peak
frequency sweep. Frequency returned is in Hz.
Query Example
On Reset
On *RST, the value is set to 50 MHz.
Error Messages
• If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
• If parameter set is lower than 1 kHz, error –222, “Data out of range;
value clipped to lower limit” occurs.
• If parameter set is higher than 1000 GHz, error –222, “Data out of
range; value clipped to upper limit” occurs.
FREQ:STOP 1MHz This command sets frequency sweep to
stop at 1 MHz for Channel A.
FREQ:STOP? This query returns the stop frequency of
frequency sweep in Hz for Channel A.