Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 23
Using READ?
Some fine tuning of the above program segment can be carried out for
example, by setting the trigger delay off. The following program segment
assumes that Channel A is currently being measured on the upper
window.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 READ1?
Using INITiate and FETCh?
The following program segment assumes that Channel A is currently being
measured on the upper window.
Some fine tuning of the above program segment can be carried out for
example, by setting the trigger delay off.
1 ABOR1
2 CONF1 DEF,3
ABOR1 Aborts Channel A
CONF1 DEF,3 Configures the upper window to make a measurement
using the current setting of the expected power and
source list and a resolution setting of 3
READ1? Takes the upper window’s measurement. This is
Channel A or B measurement depending on current
window setup.
ABOR1 Aborts Channel A
CONF1 DEF,3 Configures the upper window to make a
measurement using the current setting of the
expected power and source list and a resolution
setting of 3
INIT1 Causes Channel A to make a measurement
FETC1? DEF,3 Retrieves the upper window’s measurement