10.2 Error codes and countermeasures
Chapter 10
Error code Displayed message Problems/countermeasures
07143 - 07152
Error was detected while processing the cassette.
Please open the door and remove the Cassette and
Plate. After removal, erase the cassette in ERASE
MODE.
(1) Remove the cassette and plate from the device.
(2) Delete the image on the removed plate using the
Erase mode.
07153
Error was detected while processing the cassette.
Please open the door and remove the Cassette and
Plate. After removal, erase the cassette in ERASE
MODE.
While processing the cassette, an error was detected.
Open the door, remove the cassette/plate, and erase
the cassette using Erase mode.
07154
Error was detected while processing the cassette.
Please open the door and remove the Cassette and
Plate. After removal, erase the cassette in ERASE
MODE. Moreover, please clean a roller.
While processing the cassette, an error was detected.
Open the door, remove the cassette/plate, and erase
the cassette using Erase mode. Also, clean the rollers.
07200 - 07224
07300 - 07354
Single unit operation error. Contact Konica Minolta technical representatives.
07400 - 07409
Initialization failed. Reader error was detected during
initialization.
Open and close the top door. Perform initialization of
the device again.
Check if the error reoccurs.
07410
07411
Erasure error was detected.
An error was detected in the device when erasing the
image.
(1) As the image reading is complete, wait for the im-
age transfer to nish.
(2) After conrming that the transfer is complete,
remove the cassette.
(3) Contact Konica Minolta technical representatives.
07412 - 07414 Reader error was detected.
(1) Remove the cassette and plate from the device.
(2) Contact Konica Minolta technical representatives.
07415
Initialization failed. Reader error was detected during
initialization.
Open and close the top door. Perform initialization of
the device again.
Check if the error reoccurs.
07416 - 07419 Reader error was detected.
(1) Remove the cassette and plate from the device.
(2) Contact Konica Minolta technical representatives.
07420
Initialization failed. Reader error was detected during
initialization.
Open and close the top door. Perform initialization of
the device again.
Check if the error reoccurs.
07421
07422
Reader error was detected.
(1) Remove the cassette and plate from the device.
(2) Contact Konica Minolta technical representatives.
07600 - 07622 Single unit operation error. Contact Konica Minolta technical representatives.
07650
07651
Reader error was detected. Contact Konica Minolta technical representatives.
07700
07701
Initialization failed. Reader error was detected during
initialization.
Open and close the top door. Perform initialization of
the device again.
Check if the error reoccurs.
07710 - 07746 Detected Memory error.
After this error occurs, even if the power is turned OFF
and ON, the device error will not be cleared. Contact
Konica Minolta technical representatives.
07747 - 07749 Detected Data error. Contact Konica Minolta technical representatives.
07750 - 07759 Detected Data error.
An error was detected in the transferred data.
(1) Check the transmitted data and the communication
cable.
(2) Afterward, power the device OFF and ON.
07800
07801
Parameters which cannot be processed in the Reader
Settings are specied
An error was detected in the cassette size parameter.
(1) Remove the cassette and plate from the device.
(2) Contact Konica Minolta technical representatives.
07802 Reader error was detected.
(1) Remove the cassette and plate from the device.
(2) Contact Konica Minolta technical representatives.
07803
Parameters which cannot be processed in the Reader
Settings are specied
Exposure conditions which cannot be processed by
the device have been indicated. Recheck the expo-
sure conditions.