2 MEASurement Commands
136 N1911A/1912A P-Series Power Meters Programming Guide
1
The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the
command sub-systems. The parameters must be entered in the specified order. If parameters are
omitted, they default from the right. The parameter DEFault is used as a place holder. Specifying
DEF leaves the parameter value unchanged.
2
When the measurement result is linear this parameter represents the number of significant
digits. When the measurement result is logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001
respectively.
3
N1912A only.
Example
Error Messages
• INITiate:CONTinuous must be set to OFF on both channels, otherwise
error –213, “INIT ignored” occurs.
•If TRIGger:SOURce is set to BUS or HOLD on either channel, error –214,
“Trigger deadlock” occurs.
resolution A numeric value for the resolution. If it is
unspecified the current resolution setting is used.
If a value is entered it should correspond to the
current resolution setting otherwise an error
occurs.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1
source list This channel list specifies the channels used to
calculate the difference.
If unspecified and the current window setup is a
difference measurement then this difference
setup is used, otherwise it defaults to Channel
A-B (N1912A) or A-A (N1911A).
(@1),(@2)
3
(@2),(@1)
3
(@1),(@1)
(@2),(@2)
3
Item Description/Default Range of Values
READ2:DIFF? This command queries difference
measurement on the lower window/upper
measurement.