2 MEASurement Commands
140 N1911A/1912A P-Series Power Meters Programming Guide
Example
Error Messages
• INITiate:CONTinuous must be set to OFF on both channels, otherwise
error –213, “INIT ignored” occurs.
•If TRIGger:SOURce is set to BUS or HOLD on either channel, error
–214, “Trigger deadlock” occurs.
• If the resolution parameter is not the same as the current resolution
setting on the specified window, error –221, “Settings conflict” occurs.
READ1:DIFF:REL? DEF,4,(@2),(@1) This command queries the upper
window/upper measurement
relative difference measurement of
Channel B - Channel A, using the
current sensor range and a
resolution setting of 4 on both
channels.