CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 207
Syntax
Example
Reset Condition
On reset, automatic calibration is disabled.
Query
CALibration[1]|2:AUTO?
The query always returns a value of 0.
Error Messages
If this command is set to ON and an 8480 Series, E- Series, N8480 or
:AUTO
0|OFF
1|ON
Space
?
CAL
1
2
ONCE
CAL1:AUTO ONCE This command causes the power meter to
perform a calibration on Channel A.
NOTE
If the command is set to ON when a N1920 is connected, auto cal is enabled.