5DISPlay Subsystem
246 N1911A/1912A P-Series Power Meters Programming Guide
Query
DISPlay[:WINDow[1]|2]:STATe?
This enters a 1 or 0 in the output buffer indicating the selected window.
1 is returned if the window is enabled
0 is returned if the window is disabled
Query Example
DISP:WIND2:STAT? This command queries whether or not the
lower window is displayed.