EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 285
Syntax
Parameters
Example
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of calibration factor/offset points for the currently
selected table.
Item Description/Default Range of Values
numeric_value A numeric value for the calibration/ offset
factors. The units are PCT.
1.0 to 150.0
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
This command enters a reference
calibration factor of 97 % and calibration
factors of 99.5 % and 97.4 % into the
sensor calibration table.
MEM
:TABL
:GAIN
:MAGN
numeric_value
Space
,
?

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals