MEMory Subsystem 7
N1911A/1912A P-Series Power Meters Programming Guide 285
Syntax
Parameters
Example
Query
MEMory:TABLe:GAIN[:MAGNitude]?
The query returns a list of calibration factor/offset points for the currently
selected table.
Item Description/Default Range of Values
numeric_value A numeric value for the calibration/ offset
factors. The units are PCT.
1.0 to 150.0
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
This command enters a reference
calibration factor of 97 % and calibration
factors of 99.5 % and 97.4 % into the
sensor calibration table.
MEM
:TABL
:GAIN
:MAGN
numeric_value
Space
,
?