8 OUTput Subsystem
304 N1911A/1912A P-Series Power Meters Programming Guide
Query
OUTPut:TRIGger[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating
whether or not the trigger output signal is enabled or disabled.
1 is returned when the trigger output signal is enabled
0 is returned when the trigger output signal is disabled
Query Example
Error Messages
If 8480, N8480, E4410, E9300 or E9320 sensor is connected and the
trigger source is not set to external, error –221 “Settings conflict”
occurs.
OUTP:TRIG:STAT? This command queries the status of the
trigger output signal.