PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 331
Query
PStatistic[1]|2:CCDF CONTinuous?
The query enters 1 or 0 into the output buffer indicating the status of the
CCDF Continuous Refresh mode.
1 is returned when the CCDF Continuous Refresh mode is enabled
0 is returned when the CCDF Continuous Refresh mode is disabled (or
CCDF Single Refresh mode is enabled)
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in free- run acquisition mode, error –221 "Settings
conflict" occurs.
PST1:CCDF:CONT? This command queries whether the CCDF
Continuous Refresh mode is on or off for
Channel A.