10 SENSe Subsystem
380 N1911A/1912A P-Series Power Meters Programming Guide
Error Messages
If the command is used when a sensor other than a P- Series or E9320
power sensor is connected, error –241, “Hardware missing” occurs
If the command is used when a P- Series or an E9320 sensor is
connected and set to AVERage mode rather than NORMal mode, the
error –221, “Settings Conflict” occurs.
NOTE
Selection of video bandwidth to LOW, MED or HIGH implements digital signal processing
to ensure a flat bandwidth up to the frequency shown, bandwidths are flat to ±0.1 dB. In the
OFF state no corrections are applied and the response has a slow roll-off.