SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 379
1
At 3.0 dB roll off point.
Example
Reset Condition
On reset, sensor bandwidth is set to OFF.
Query
[SENSe[1]]|SENSe2:BANDwidth|BWIDth:VIDeo?
The query returns the current sensor bandwidth setting.
Query Example
N1920A
N1921A
5 MHz 15 MHz 30 MHz 30 MHz
Video Bandwidth Settings
Sensor LOW MEDium HIGH OFF
SENSe1:BAND:VID HIGH This command sets sensor bandwidth to
high for Channel A.
SENS2:BAND:VID? This command queries the current sensor
bandwidth setting for Channel B.