EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 379
1
At 3.0 dB roll off point.
Example
Reset Condition
On reset, sensor bandwidth is set to OFF.
Query
[SENSe[1]]|SENSe2:BANDwidth|BWIDth:VIDeo?
The query returns the current sensor bandwidth setting.
Query Example
N1920A
N1921A
5 MHz 15 MHz 30 MHz 30 MHz
Video Bandwidth Settings
Sensor LOW MEDium HIGH OFF
SENSe1:BAND:VID HIGH This command sets sensor bandwidth to
high for Channel A.
SENS2:BAND:VID? This command queries the current sensor
bandwidth setting for Channel B.

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals