10 SENSe Subsystem
382 N1911A/1912A P-Series Power Meters Programming Guide
Example
Query
[SENSe[1]]|SENSe2:BUFFer:COUNt?
This query is used to retrieve the average or peak measurement buffer
size.
Query Example
On Reset
On *RST, the value is set to 1.
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241,
“Hardware missing” occurs.
If E4410, N8480, E9300, E9320 or N1920 sensor is connected but
acquisition mode is in free run, error –221, “Setting conflict. Invalid
acquisition mode” occurs.
If frequency sweep step is non- zero, error –221, “Settings conflict.
Frequency sweep enabled. Buffer count overidden” occurs.
BUFF:COUN 100 This command sets the average or peak
measurement buffer size to 100 for
Channel A.
BUFF:COUN? This query returns the average or peak
measurement buffer size for Channel A.