10 SENSe Subsystem
402 N1911A/1912A P-Series Power Meters Programming Guide
• 1 is returned when the pulse power measurement feature is enabled
• 0 is returned when the pulse power measurement feature is disabled
Query Example
Error Messages
•If [SENSe[1]]|SENSe2:CORRection:DCYCle:STATus is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings
Conflict” occurs.
• If this command is used when an E4412A/E4413A power sensor is
connected, the error –310, “System error;Dty Cyc may impair accuracy
with ECP sensor” occurs. If you are using a dual channel power meter
the error message specifies the channel.
• If this command is used when a E9320 power sensor is set to NORMal
mode, the error –221, “Settings Conflict” occurs.
• If this command is used when a P- Series power sensor is connected,
the error –241, “Settings Conflict” occurs.
CORR:GAIN3:STAT? This command queries whether the pulse
measurement feature is on or off.