EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 SENSe Subsystem
406 N1911A/1912A P-Series Power Meters Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the
channel offsets.
1 is returned if a channel offset is enabled
0 is returned if a channel offset is disabled
Query Example
Error Messages
If [SENSe[1]]|SENSe2:CORRection:GAIN2:STATe is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings
Conflict” occurs.
CORR:GAIN2:STAT? This command queries whether or not
there is a channel offset applied to
Channel A.

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals