10 SENSe Subsystem
406 N1911A/1912A P-Series Power Meters Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe?
The query enters 1 or 0 into the output buffer indicating the status of the
channel offsets.
1 is returned if a channel offset is enabled
0 is returned if a channel offset is disabled
Query Example
Error Messages
If [SENSe[1]]|SENSe2:CORRection:GAIN2:STATe is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings
Conflict” occurs.
CORR:GAIN2:STAT? This command queries whether or not
there is a channel offset applied to
Channel A.