SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 413
Parameters
1
The following measurement units can be used:
•Hz
• kHz (10
3
)
• MHz (10
6
)
• GHz (10
9
)
Example
Reset Condition
On reset, the frequency is set to 50 MHz (DEF).
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]? [MIN|MAX]
The query returns the current frequency setting or the values associated
with MIN and MAX. The units in which the results are returned are Hz.
Item Description/Default Range of Values
numeric_value A numeric value for the frequency:
• DEF: the default value is 50 MHz
• MIN: 1 kHz
• MAX: 1000.0 GHz
The default units are Hz.
1 kHz to 1000.0 Ghz
1
DEF
MIN
MAX
FREQ 500kHz This command enters a Channel A
frequency of 500 kHz.