SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 441
Reset Condition
On reset, gate 1 is set to 100 µs and other gates to 0 s.
Query
SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME?
The query returns the current length of the time- gated period.
Query Example
Error Messages
If the command is used when a sensor other than a P- Series or E9320
power sensor is connected, error –241, “Hardware missing” occurs
If the command is used when an E9320 sensor is connected and set to
AVERage mode rather than NORMal mode, the error –221, “Settings
Conflict” occurs.
SENS2:SWE2:TIME? This command queries the length of the
time- gated period for Channel B and gate
2.