10 SENSe Subsystem
456 N1911A/1912A P-Series Power Meters Programming Guide
Parameters
Example
Reset Condition
On reset, the value is set to –50 dBm.
Query
SENSe[1]|2:TRACe:LIMit:LOWer [MIN|MAX]
The query returns the current setting of the trace lower scale limit or the
value associated with MIN or MAX. The format of the response is <NR3>.
Item Description/Default Range of Values
numeric_value A numeric value for the trace lower scale
limit.
• DEF: the default is 20 dBm
• MIN: –150 dBm
• MAX: 230 dBm
–150 to 230 dBm
DEF
MIN
MAX
SENS2:TRAC:LIM:LOW 10 This command sets the trace lower scale
limit to 10 dBm for Channel B.