Power Meter Remote Operation 1
N1911A/1912A P-Series Power Meters Programming Guide 21
Using READ?
Some fine tuning of measurements can be performed using the CONFigure
and READ? commands. For example, in the above program segment some
fine tuning can be performed by setting the filter length to 1024 and the
trigger delay off.
1 ABOR2
2 CONF1 -50,DEF,(@2)
3 SENS2:AVER:COUN 1024
4 TRIG2:DEL:AUTO OFF
5 READ1?
Using INITiate and FETCh?
Some fine tuning of measurements can be carried out using the
CONFigure command and INITiate and FETCh? commands. For example,
in the above program segment some fine tuning can be carried out by
ABOR2 Aborts Channel B
CONF1 -50,DEF,(@2) Configures the upper window to make a
Channel B measurement using an expected
power of –50 dBm and the current resolution
setting
READ1? Takes the upper window’s measurement
ABOR2 Aborts Channel B
CONF1 -50,DEF,(@2) Configures the upper window to make a
Channel B measurement using an expected
power of –50 dBm and the current resolution
setting
INIT2 Causes Channel B to make a measurement
FETC1? -50,DEF,(@2) Retrieves the upper window’s measurement

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals