STATus Subsystem 11
N1911A/1912A P-Series Power Meters Programming Guide 469
[:EVENt]?
This query returns a 16 bit decimal- weighted number representing the bits
set in the Event Register of the SCPI Register Set you require to control.
The format of the return is <NR1> in the range of 0 to 32767 (2
15
–1). This
query clears all bits in the register to 0.
Syntax
:ENABle <NRf>|<non-decimal numeric>
This command sets the Enable Register of the particular SCPI Register Set
you require to control. The parameter value, when rounded to an integer
and expressed in base 2 has its first 15 bits written into the Enable
Register of the SCPI Register Set concerned. The last bit (bit 15) is always
set to 0.
Syntax
NOTE
The [:EVENt]? is the default command if the STATus SCPI are not accompanied by any of
the Status Register Set commands (:COND, :ENAB, :NTR and :PTR).
:EVEN
?
:ENAB
NRf
non-decimal numeric
?