EasyManuals Logo

Agilent Technologies N1911A Programming Guide

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #645 background imageLoading...
Page #645 background image
TRACe Subsystem 13
N1911A/1912A P-Series Power Meters Programming Guide 617
Example
Reset Condition
On reset, the reference level will set to 10 % and 90 % respectively.
Query
TRACe[1]|2:DEFine:TRANsition:REFerence?
The query returns trace 1 reference levels used in the transition
occurences calculation.
Query Example
TRAC1:DEF:TRAN:REF 1,18 This command sets trace 1 transition
measurements to look for the 1 % and 81 %
reference levels.
TRAC1:DEF:TRAN:REF
DEF,DEF
This command sets trace 1 transition
measuremetns to look for the 10 % and 90
% reference levels.
TRAC1:DEF:TRAN:REF? This command queries the reference levels
used in the calculation of transition
durations and occurrences for trace 1.

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals