EasyManua.ls Logo

RFL Electronics GARD 8000 - Page 441

RFL Electronics GARD 8000
647 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Tests and Diagnostics
GARD 8000 SYS RFL Electronics
February 14, 2011 11-7 973.334.3100
The page consists of three sections, status, test buttons and test configuration.
The status section consists of the following read only fields.
Start of Logic Range:
Indicates the starting logic bit and if the module is using the default logic assignment or custom logic
assignment.
Description:
Each of the 6 output functions include the description programmed in with the GARD 8000 logic file
or in the output module configuration screen.
Slot #:
Indicates the physical slot number of the module in the GARD 8000 chassis.
TB#:
Gives the terminal block numbers of the output function.
State:
Provides the logic state of the function when the page is loaded or refreshed.
Bit assignment:
Indicates the logic bit the output function is driven by on the logic bus.
Enabled?:
Indicates if the output function is enabled on the Input settings page. If the output is enabled the
GARD 8000 logic bus will drive that output function.
There are five buttons associated with the Output Module test page. They are:
Module Enable:
This enables the output module. The GARD 8000 logic bus drives the output functions.
Module Disable
This disables the Output module. All six of the output functions of that module are blocked
from the GARD 8000 logic bus. Changes to GARD 8000 logic bus will not change the output
function.
Set all to Normal
Returns the mode of all the output functions to normal operation. Normal operation will result
in the output being driven by the logic bus. This function will be used to return the module to
normal operating condition after performing tests that require changes to the output mode as
described below.
Execute test
When the Execute test button is pressed the state of the 6 output functions will be changed to
the state described in the Mode section below.
All output functions can be returned to normal state by setting the function mode to normal on
all functions and hitting the execute test button or by hitting the Set All to Normal button
described above.

Table of Contents