SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 367
CONFigure:POWer:AC? commands.
Syntax
Example
Reset Condition
On reset, automatic averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO?
NOTE
For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best mode of operation.
However, manual filter length selection ([SENSe[1]]|SENSe2:AVERage:COUNt
<numeric_value>) is useful in applications requiring either high resolution or fast
settling times, where signal variations rather than measurement noise need filtering, or
when approximate results are needed quickly.
AVER:COUN:AUTO OFF This command disables automatic filter
length selection for Channel A.
SENS
1
:COUN
AVER 0|OFF
1|ON
Space
?
SENS2
:AUTO
:
ONCE