Startup
106
Instructions for use Zeus Infinity Empowered SW 2.n
Canceling the self test
z Touch the Cancel button (C). The button turns
yellow. Confirm with rotary knob.
No message appears on the system standby
screen. The Startup dialog dialog window will be
opened.
The Ventilator MAN/SPON button (D) is shown in
yellow. If the Startup dialog is confirmed with the
rotary know, the devices switches automatically to
the Ventilator MAN/SPON mode with the preset
profile.
z Touch the MAN/SPON field (F).
The Ventilation settings dialog window opens.
1 Touch the button for the required ventilation
mode.
2 Adjust the fresh-gas settings if necessary.
3 Return the safety rotary knob O
2 to 0 and push
in.
312323
Startup dialog
D
MAN/SPON.
F
NOTE
The warm-up phase of the DIVA metering mod-
ules takes approx. 2 minutes. No anesthetic gas
can be delivered during this time!
The leak and compliance test is not performed.
The levels of accuracy indicated under "Technical
data" (e.g. flow measurement) cannot be guaran-
teed.
WARNING
In order to ensure device features in an emer-
gency operation, check APL valve and breath-
ing system operation by ventilating manually
in MAN/SPON ventilation mode.