Alarms
182
Instructions for use Zeus Infinity Empowered SW 2.n
Arrhythmia detection
T1/T2 (°C) –5 to 49.9 34
TBlood (°C) 25.1 to 43 39
TBlood (°C) 25 to 42.9 34
TOF-Count 1 to 4; Off 4
TOF-Count 0 to 3; Off 0
BIS 15 to 100; Off 100
BIS 0 to 95; Off 10
Alarm Setting range Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1
Alarm Setting range Resolu-
tion/incre-
ment
Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1
ASY
Alarm On/Off OnOnOnOn Off
Rate – – – – – –
Count – – – – – –
VF
Alarm On/Off OnOnOnOn Off
Rate – – – – – –
Count – – – – – –
V.tach
Alarm On/Off OnOnOnOn Off
Rate 100 to 200 10 ≥120 ≥120 ≥120 ≥120 –
Count 5 to 15 1 ≥10 ≥10 ≥10 ≥10 –
V.run
Alarm On/Off OnOnOnOn Off
Rate 100 to 200 10 ≥120 ≥120 ≥120 ≥120 –
Count 3 to (VTcount –1) 1 3 to 9 3 to 9 3 to 9 3 to 9 –
AIVR
Alarm On/Off OnOnOnOn Off
Rate 100 to 200
(<VTcount –1)
10 119 119 119 119 –
Count count >3 3 3 3 3 –