EasyManua.ls Logo

Dräger Zeus Infinity Empowered - Arrhythmia Detection

Dräger Zeus Infinity Empowered
453 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
182
Instructions for use Zeus Infinity Empowered SW 2.n
Arrhythmia detection
T1/T2 (°C) 5 to 49.9 34
TBlood (°C) 25.1 to 43 39
TBlood (°C) 25 to 42.9 34
TOF-Count 1 to 4; Off 4
TOF-Count 0 to 3; Off 0
BIS 15 to 100; Off 100
BIS 0 to 95; Off 10
Alarm Setting range Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1
Alarm Setting range Resolu-
tion/incre-
ment
Default value of factory profile
Adult 2 Adult 1 Ped 2 Ped 1 Neo 1
ASY
Alarm On/Off OnOnOnOn Off
Rate
Count
VF
Alarm On/Off OnOnOnOn Off
Rate
Count
V.tach
Alarm On/Off OnOnOnOn Off
Rate 100 to 200 10 120 120 120 120
Count 5 to 15 1 10 10 10 10
V.run
Alarm On/Off OnOnOnOn Off
Rate 100 to 200 10 120 120 120 120
Count 3 to (VTcount –1) 1 3 to 9 3 to 9 3 to 9 3 to 9
AIVR
Alarm On/Off OnOnOnOn Off
Rate 100 to 200
(<VTcount –1)
10 119 119 119 119
Count count >3 3 3 3 3

Table of Contents

Related product manuals