Cleaning, disinfection, and sterilization
374
Instructions for use Zeus Infinity Empowered SW 2.n
Surfaces of Zeus IE, pressurized gas
hoses, cables, and DIVA metering mod-
ules
Screen of Zeus IE
Endotracheal suction After each pa-
tient
Observe the associated instructions for use.
SpiroLife flow sensor Weekly Observe the associated instructions for use.
Items which can be re-
processed
Recommended
reprocessing in-
tervals
Prelimi-
nary
cleaning
Machine
cleaning and
disinfection
Manual Sterilization
Cleaning Disinfec-
tion
WARNING
Do not allow any liquids to get into the device
openings! This can damage the device. Do not
treat the surface of the device with products
containing alcohol or else the surface may be
damaged!
WARNING
Do not allow any detergents or substances not
intended for use to enter the DIVA metering
module. The anesthetic agent may be contam-
inated. Risk of patient injury.
WARNING
Due to a risk of electric shock, the device
cover may not be removed if in operation or a
power supply is connected.
CAUTION
Do not used sharp tools or abrasives. Never put
electrical connections in water or other liquids.
CAUTION
Do not put the screen and its accessories in liquid
or rinse with liquids. If liquids accidentally get on
the device, immediately unplug the device from
the power supply. Contact service personnel to
obtain information about further device safety be-
fore putting the device back into operation.
NOTE
Never clean the front of the screen while monitor-
ing a patient. Put the device out of operation be-
fore cleaning.
CAUTION
When using alcohol, a solution diluted to 40 %
should be used. Higher concentrations can dam-
age the device.
CAUTION
Do not spray disinfectant on the touch screen di-
rectly. The cleaning agent must be transferred to
a cloth or towel before cleaning the surface.